A system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization

Electron capture negative ion mass spectroscopy (ECNIMS) has been performed on environmental chemicals and explosives. A trochoidal electron monochromator interfaced to a gas chromatograph and a quadrupole mass spectrometer allows compounds to be studied by this method. The method employed here diff...

Full description

Bibliographic Details
Main Author: Sonnen, Steven J.
Other Authors: Kocher, Carl A., Physics, Oregon State University. Graduate School
Format: Master Thesis
Language:English
unknown
Published: Oregon State University
Subjects:
Online Access:https://ir.library.oregonstate.edu/concern/graduate_thesis_or_dissertations/4j03d2667
id ftoregonstate:ir.library.oregonstate.edu:4j03d2667
record_format openpolar
spelling ftoregonstate:ir.library.oregonstate.edu:4j03d2667 2024-09-09T19:24:57+00:00 A system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization Sonnen, Steven J. Kocher, Carl A. Physics Oregon State University. Graduate School https://ir.library.oregonstate.edu/concern/graduate_thesis_or_dissertations/4j03d2667 English [eng] eng unknown Oregon State University https://ir.library.oregonstate.edu/concern/graduate_thesis_or_dissertations/4j03d2667 Copyright Not Evaluated Explosives -- Testing Terrorism -- Prevention Masters Thesis ftoregonstate 2024-07-22T18:06:05Z Electron capture negative ion mass spectroscopy (ECNIMS) has been performed on environmental chemicals and explosives. A trochoidal electron monochromator interfaced to a gas chromatograph and a quadrupole mass spectrometer allows compounds to be studied by this method. The method employed here differs from standard ECNIMS in that no reagent gas is used to moderate electron energies. Several explosives were analyzed using this system, as were chlorinated compounds obtained from the lipid fraction of Arctic trout muscle. A miniaturized version of the system would be advantageous as a portable apparatus for field use. Among the required modifications would be the installation of cylindrical permanent magnets to replace the Helmholtz coils used for generating the magnetic field for the monochromator. Tests suggest that the trochoidal electron monochromator component of the system can be reduced in volume by a factor of 200 without an appreciable loss of energy resolution. Master Thesis Arctic ScholarsArchive@OSU (Oregon State University) Arctic
institution Open Polar
collection ScholarsArchive@OSU (Oregon State University)
op_collection_id ftoregonstate
language English
unknown
topic Explosives -- Testing
Terrorism -- Prevention
spellingShingle Explosives -- Testing
Terrorism -- Prevention
Sonnen, Steven J.
A system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization
topic_facet Explosives -- Testing
Terrorism -- Prevention
description Electron capture negative ion mass spectroscopy (ECNIMS) has been performed on environmental chemicals and explosives. A trochoidal electron monochromator interfaced to a gas chromatograph and a quadrupole mass spectrometer allows compounds to be studied by this method. The method employed here differs from standard ECNIMS in that no reagent gas is used to moderate electron energies. Several explosives were analyzed using this system, as were chlorinated compounds obtained from the lipid fraction of Arctic trout muscle. A miniaturized version of the system would be advantageous as a portable apparatus for field use. Among the required modifications would be the installation of cylindrical permanent magnets to replace the Helmholtz coils used for generating the magnetic field for the monochromator. Tests suggest that the trochoidal electron monochromator component of the system can be reduced in volume by a factor of 200 without an appreciable loss of energy resolution.
author2 Kocher, Carl A.
Physics
Oregon State University. Graduate School
format Master Thesis
author Sonnen, Steven J.
author_facet Sonnen, Steven J.
author_sort Sonnen, Steven J.
title A system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization
title_short A system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization
title_full A system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization
title_fullStr A system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization
title_full_unstemmed A system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization
title_sort system for analyzing and characterizing environmental chemicals and explosives and a propoasal for its miniaturization
publisher Oregon State University
url https://ir.library.oregonstate.edu/concern/graduate_thesis_or_dissertations/4j03d2667
geographic Arctic
geographic_facet Arctic
genre Arctic
genre_facet Arctic
op_relation https://ir.library.oregonstate.edu/concern/graduate_thesis_or_dissertations/4j03d2667
op_rights Copyright Not Evaluated
_version_ 1809894768443916288