A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method

We report a high performance autonomous analytical system based on the vanadomolybdate method for the determination of soluble reactive phosphorus in seawater. The system combines a microfluidic chip manufactured from tinted poly (methyl methaaylate) (PMMA), a custom made syringe pump, embedded cont...

Full description

Bibliographic Details
Published in:Talanta
Main Authors: Legiret, François-Eric, Sieben, Vincent J., Woodward, E. Malcolm S., Abi Kaed Bey, Samer K., Mowlem, Matthew C., Connelly, Douglas P., Achterberg, Eric P.
Format: Article in Journal/Newspaper
Language:unknown
Published: Elsevier 2013
Subjects:
Online Access:https://oceanrep.geomar.de/id/eprint/22818/
https://doi.org/10.1016/j.talanta.2013.05.004
id ftoceanrep:oai:oceanrep.geomar.de:22818
record_format openpolar
spelling ftoceanrep:oai:oceanrep.geomar.de:22818 2023-05-15T17:34:29+02:00 A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method Legiret, François-Eric Sieben, Vincent J. Woodward, E. Malcolm S. Abi Kaed Bey, Samer K. Mowlem, Matthew C. Connelly, Douglas P. Achterberg, Eric P. 2013-11-15 https://oceanrep.geomar.de/id/eprint/22818/ https://doi.org/10.1016/j.talanta.2013.05.004 unknown Elsevier Legiret, F. E., Sieben, V. J., Woodward, E. M. S., Abi Kaed Bey, S. K., Mowlem, M. C., Connelly, D. P. and Achterberg, E. P. (2013) A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method. Talanta, 116 . pp. 382-387. DOI 10.1016/j.talanta.2013.05.004 <https://doi.org/10.1016/j.talanta.2013.05.004>. doi:10.1016/j.talanta.2013.05.004 Article PeerReviewed 2013 ftoceanrep https://doi.org/10.1016/j.talanta.2013.05.004 2023-04-07T15:11:28Z We report a high performance autonomous analytical system based on the vanadomolybdate method for the determination of soluble reactive phosphorus in seawater. The system combines a microfluidic chip manufactured from tinted poly (methyl methaaylate) (PMMA), a custom made syringe pump, embedded control electronics and on-board calibration standards. This "lab-on-a-chip" analytical system was successfully deployed and cross-compared with reference analytical methods in coastal (south west England) and open ocean waters (tropical North Atlantic). The results of the miniaturized system compared well with a reference bench-operated phosphate auto-analyser and showed no significant differences in the analytical results (student's t-test at 95% confidence level). The optical technology used, comprising of tinted PMMA and polished fluidic channels, has allowed an improvement of two orders of magnitude of the limit of detection (52 nM) compared to currently available portable systems based on this method. The system has a wide linear dynamic range 0.1-60 mu M, and a good precision (13.6% at 0.4 mu M, n=4). The analytical results were corrected for silicate interferences at 0.7 mu M, and the measurement frequency was configurable with a sampling throughput of up to 20 samples per hour. This portable micro-analytical system has a low reagent requirement (340 mu L per sample) and power consumption (756J per sample), and has allowed accurate high resolution measurements of soluble reactive phosphorus in seawater Article in Journal/Newspaper North Atlantic OceanRep (GEOMAR Helmholtz Centre für Ocean Research Kiel) Talanta 116 382 387
institution Open Polar
collection OceanRep (GEOMAR Helmholtz Centre für Ocean Research Kiel)
op_collection_id ftoceanrep
language unknown
description We report a high performance autonomous analytical system based on the vanadomolybdate method for the determination of soluble reactive phosphorus in seawater. The system combines a microfluidic chip manufactured from tinted poly (methyl methaaylate) (PMMA), a custom made syringe pump, embedded control electronics and on-board calibration standards. This "lab-on-a-chip" analytical system was successfully deployed and cross-compared with reference analytical methods in coastal (south west England) and open ocean waters (tropical North Atlantic). The results of the miniaturized system compared well with a reference bench-operated phosphate auto-analyser and showed no significant differences in the analytical results (student's t-test at 95% confidence level). The optical technology used, comprising of tinted PMMA and polished fluidic channels, has allowed an improvement of two orders of magnitude of the limit of detection (52 nM) compared to currently available portable systems based on this method. The system has a wide linear dynamic range 0.1-60 mu M, and a good precision (13.6% at 0.4 mu M, n=4). The analytical results were corrected for silicate interferences at 0.7 mu M, and the measurement frequency was configurable with a sampling throughput of up to 20 samples per hour. This portable micro-analytical system has a low reagent requirement (340 mu L per sample) and power consumption (756J per sample), and has allowed accurate high resolution measurements of soluble reactive phosphorus in seawater
format Article in Journal/Newspaper
author Legiret, François-Eric
Sieben, Vincent J.
Woodward, E. Malcolm S.
Abi Kaed Bey, Samer K.
Mowlem, Matthew C.
Connelly, Douglas P.
Achterberg, Eric P.
spellingShingle Legiret, François-Eric
Sieben, Vincent J.
Woodward, E. Malcolm S.
Abi Kaed Bey, Samer K.
Mowlem, Matthew C.
Connelly, Douglas P.
Achterberg, Eric P.
A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method
author_facet Legiret, François-Eric
Sieben, Vincent J.
Woodward, E. Malcolm S.
Abi Kaed Bey, Samer K.
Mowlem, Matthew C.
Connelly, Douglas P.
Achterberg, Eric P.
author_sort Legiret, François-Eric
title A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method
title_short A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method
title_full A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method
title_fullStr A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method
title_full_unstemmed A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method
title_sort high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method
publisher Elsevier
publishDate 2013
url https://oceanrep.geomar.de/id/eprint/22818/
https://doi.org/10.1016/j.talanta.2013.05.004
genre North Atlantic
genre_facet North Atlantic
op_relation Legiret, F. E., Sieben, V. J., Woodward, E. M. S., Abi Kaed Bey, S. K., Mowlem, M. C., Connelly, D. P. and Achterberg, E. P. (2013) A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method. Talanta, 116 . pp. 382-387. DOI 10.1016/j.talanta.2013.05.004 <https://doi.org/10.1016/j.talanta.2013.05.004>.
doi:10.1016/j.talanta.2013.05.004
op_doi https://doi.org/10.1016/j.talanta.2013.05.004
container_title Talanta
container_volume 116
container_start_page 382
op_container_end_page 387
_version_ 1766133330617565184