Calibration of the superconducting gravimeter CT#043 with an absolute gravimeter FG5#210 at Syowa Station, Antarctica

P(論文) For the purpose of calibration of the superconducting gravimeter (SG) model CT#043 at Syowa Station, Antarctica, we carried out parallel observations with the absolute gravimeter (AG) FG5#210. Although the AG measurements were conducted from December 31, 2003 to February 9, 2004, SG measuremen...

Full description

Bibliographic Details
Main Authors: Fukuda, Yoichi, Iwano, Sachiko, Ikeda, Hiroshi, Hiraoka, Yoshifumi, Doi, Koichiro
Language:English
Published: National Institute of Polar Research 2005
Subjects:
Online Access:https://nipr.repo.nii.ac.jp/record/3148/files/KJ00003562984.pdf
https://doi.org/10.15094/00003148
https://nipr.repo.nii.ac.jp/records/3148
Description
Summary:P(論文) For the purpose of calibration of the superconducting gravimeter (SG) model CT#043 at Syowa Station, Antarctica, we carried out parallel observations with the absolute gravimeter (AG) FG5#210. Although the AG measurements were conducted from December 31, 2003 to February 9, 2004, SG measurements before January 17 are found to have suffered artificial sensitivity change. We finally adopted data from January 17 to February 1, 2004, and obtained the scale factor of CT#043 as -59.461 ±0.079μGal/volt by linear regression analysis. The result achieved a relative accuracy of around 0.1%, which is important to modern precise tidal analysis. We also determined the scale factor by comparing the data of parallel observations with the SG TT70#016. Both values showed good agreement, supporting the reliability of the above mentioned value. departmental bulletin paper