Summary: | The Lunar Advanced Volatiles Analysis (LAVA) and the Oxygen Volatiles Extraction Node (OVEN) are subsystems included in the Regolith Environment Science, and Oxygen Lunar Volatiles Extraction (RESOLVE) payload bound for the Moon in 2019. This Resource Prospector Mission (RPM) has the objective of landing on a shadowed region of the Moons South Pole to collect data and determine whether the resources could be effectively used for space exploration systems. The quantification of the resources will help understand if it can adequately minimize materials carried from Earth by: providing life support, propellants, construction materials or energy supply to the payload or crew. This paper outlines the procedures done for the pressure analysis of the LAVA-OVEN (LOVEN) Integration Testing. The pressure analysis quantifies how much gases and water are present in the sample tested during the Engineering Testing Unit (ETU) phase of instrument development. Ultimately the purpose of these tests is to improve the estimate of the amount of water in each Lunar sample and reduce the time necessary for this estimate. The governing principle that was used for the analysis is the Ideal Gas Law, PV=nRT where P stands for pressure, V for volume, n for number of moles, R being the gas constant and T for temperature. We also estimate the errors involved in these measured and derived quantities since a key objective of the mission is to estimate the quantity of volatiles present in the lunar samples introduced into OVEN.
|