Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish

Haemoglobin concentration ([Hb]) is measured for a wide variety of animal studies. The use of point-of-care devices, such as the HemoCue, is becoming increasingly common because of their portability, relative ease of use and low cost. In this study, we aimed to determine whether the [Hb] of blue-ton...

Full description

Bibliographic Details
Published in:Conservation Physiology
Main Authors: Andrewartha, Sarah J., Munns, Suzanne L., Edwards, Ashley
Format: Article in Journal/Newspaper
Language:unknown
Published: Oxford University Press 2016
Subjects:
Online Access:https://researchonline.jcu.edu.au/43423/1/Conserv%20Physiol-2016-Andrewartha-.pdf
id ftjamescook:oai:researchonline.jcu.edu.au:43423
record_format openpolar
spelling ftjamescook:oai:researchonline.jcu.edu.au:43423 2024-02-11T10:02:06+01:00 Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish Andrewartha, Sarah J. Munns, Suzanne L. Edwards, Ashley 2016 application/pdf https://researchonline.jcu.edu.au/43423/1/Conserv%20Physiol-2016-Andrewartha-.pdf unknown Oxford University Press http://dx.doi.org/10.1093/conphys/cow006 https://researchonline.jcu.edu.au/43423/ https://researchonline.jcu.edu.au/43423/1/Conserv%20Physiol-2016-Andrewartha-.pdf Andrewartha, Sarah J., Munns, Suzanne L., and Edwards, Ashley (2016) Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish. Conservation Physiology, 4 (1). cow006. pp. 1-6. open Article PeerReviewed 2016 ftjamescook https://doi.org/10.1093/conphys/cow006 2024-01-22T23:37:28Z Haemoglobin concentration ([Hb]) is measured for a wide variety of animal studies. The use of point-of-care devices, such as the HemoCue, is becoming increasingly common because of their portability, relative ease of use and low cost. In this study, we aimed to determine whether the [Hb] of blue-tongued skink (Tiliqua nigrolutea) blood can be determined accurately using the HemoCue and whether the HemoCue overestimates the [Hb] of reptile blood in a similar manner to fish blood. Additionally, we aimed to test whether ploidy affected [Hb] determined by the HemoCue using blood from diploid and triploid Atlantic salmon (Salmo salar). The HemoCue Hb 201⁺ systematically overestimated [Hb] in both blue-tongued skinks and Atlantic salmon, and there was no difference between calibration equations determined for diploid or triploid salmon. The overestimation was systematic in both species and, as such, [Hb] determined by the HemoCue can be corrected using appropriate calibration equations. Article in Journal/Newspaper Atlantic salmon Salmo salar James Cook University, Australia: ResearchOnline@JCU Conservation Physiology 4 1 cow006
institution Open Polar
collection James Cook University, Australia: ResearchOnline@JCU
op_collection_id ftjamescook
language unknown
description Haemoglobin concentration ([Hb]) is measured for a wide variety of animal studies. The use of point-of-care devices, such as the HemoCue, is becoming increasingly common because of their portability, relative ease of use and low cost. In this study, we aimed to determine whether the [Hb] of blue-tongued skink (Tiliqua nigrolutea) blood can be determined accurately using the HemoCue and whether the HemoCue overestimates the [Hb] of reptile blood in a similar manner to fish blood. Additionally, we aimed to test whether ploidy affected [Hb] determined by the HemoCue using blood from diploid and triploid Atlantic salmon (Salmo salar). The HemoCue Hb 201⁺ systematically overestimated [Hb] in both blue-tongued skinks and Atlantic salmon, and there was no difference between calibration equations determined for diploid or triploid salmon. The overestimation was systematic in both species and, as such, [Hb] determined by the HemoCue can be corrected using appropriate calibration equations.
format Article in Journal/Newspaper
author Andrewartha, Sarah J.
Munns, Suzanne L.
Edwards, Ashley
spellingShingle Andrewartha, Sarah J.
Munns, Suzanne L.
Edwards, Ashley
Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
author_facet Andrewartha, Sarah J.
Munns, Suzanne L.
Edwards, Ashley
author_sort Andrewartha, Sarah J.
title Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
title_short Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
title_full Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
title_fullStr Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
title_full_unstemmed Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
title_sort calibration of the hemocue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
publisher Oxford University Press
publishDate 2016
url https://researchonline.jcu.edu.au/43423/1/Conserv%20Physiol-2016-Andrewartha-.pdf
genre Atlantic salmon
Salmo salar
genre_facet Atlantic salmon
Salmo salar
op_relation http://dx.doi.org/10.1093/conphys/cow006
https://researchonline.jcu.edu.au/43423/
https://researchonline.jcu.edu.au/43423/1/Conserv%20Physiol-2016-Andrewartha-.pdf
Andrewartha, Sarah J., Munns, Suzanne L., and Edwards, Ashley (2016) Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish. Conservation Physiology, 4 (1). cow006. pp. 1-6.
op_rights open
op_doi https://doi.org/10.1093/conphys/cow006
container_title Conservation Physiology
container_volume 4
container_issue 1
container_start_page cow006
_version_ 1790597993852829696