Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation

International audience BACKGROUND: Compared with the first-generation Arctic Front cryoballoon (ARC-CB), the new Arctic Front Advance cryoballoon (ARC-Adv-CB) increases the efficient CB-tissue contact surface during freezing, which may increase the incidence of phrenic nerve (PN) palsy (PNP). OBJECT...

Full description

Bibliographic Details
Published in:Heart Rhythm
Main Authors: Martins, Raphaël, Hamon, David, Césari, Olivier, Behaghel, Albin, Behar, Nathalie, Sellal, Jean-Marc, Daubert, Jean-Claude, Mabo, Philippe, Pavin, Dominique
Other Authors: Service de cardiologie et maladies vasculaires Rennes = Cardiac, Thoracic, and Vascular Surgery Rennes, Centre Hospitalier Universitaire de Rennes CHU Rennes = Rennes University Hospital Pontchaillou, Laboratoire Traitement du Signal et de l'Image (LTSI), Université de Rennes (UR)-Institut National de la Santé et de la Recherche Médicale (INSERM), CIC-IT Rennes, Centre Hospitalier Universitaire de Rennes CHU Rennes = Rennes University Hospital Pontchaillou -Institut National de la Santé et de la Recherche Médicale (INSERM), Clinique Saint Gatien
Format: Article in Journal/Newspaper
Language:English
Published: HAL CCSD 2014
Subjects:
Online Access:https://hal.univ-lorraine.fr/hal-03233143
https://doi.org/10.1016/j.hrthm.2014.01.002
id ftinserm:oai:HAL:hal-03233143v1
record_format openpolar
spelling ftinserm:oai:HAL:hal-03233143v1 2024-09-15T18:11:07+00:00 Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation Martins, Raphaël Hamon, David Césari, Olivier Behaghel, Albin Behar, Nathalie Sellal, Jean-Marc Daubert, Jean-Claude Mabo, Philippe Pavin, Dominique Service de cardiologie et maladies vasculaires Rennes = Cardiac, Thoracic, and Vascular Surgery Rennes Centre Hospitalier Universitaire de Rennes CHU Rennes = Rennes University Hospital Pontchaillou Laboratoire Traitement du Signal et de l'Image (LTSI) Université de Rennes (UR)-Institut National de la Santé et de la Recherche Médicale (INSERM) CIC-IT Rennes Centre Hospitalier Universitaire de Rennes CHU Rennes = Rennes University Hospital Pontchaillou -Institut National de la Santé et de la Recherche Médicale (INSERM) Clinique Saint Gatien 2014-03 https://hal.univ-lorraine.fr/hal-03233143 https://doi.org/10.1016/j.hrthm.2014.01.002 en eng HAL CCSD Elsevier info:eu-repo/semantics/altIdentifier/doi/10.1016/j.hrthm.2014.01.002 info:eu-repo/semantics/altIdentifier/pmid/24389575 hal-03233143 https://hal.univ-lorraine.fr/hal-03233143 doi:10.1016/j.hrthm.2014.01.002 PUBMED: 24389575 ISSN: 1556-3871 EISSN: 1547-5271 Heart Rhythm https://hal.univ-lorraine.fr/hal-03233143 Heart Rhythm, 2014, 11 (3), pp.386-393. &#x27E8;10.1016/j.hrthm.2014.01.002&#x27E9; Cryoablation Paroxysmal atrial fibrillation Cryoballoon ablation Pulmonary vein Pulmonary vein isolation Phrenic nerve palsy [SDV.MHEP]Life Sciences [q-bio]/Human health and pathology info:eu-repo/semantics/article Journal articles 2014 ftinserm https://doi.org/10.1016/j.hrthm.2014.01.002 2024-07-22T23:41:41Z International audience BACKGROUND: Compared with the first-generation Arctic Front cryoballoon (ARC-CB), the new Arctic Front Advance cryoballoon (ARC-Adv-CB) increases the efficient CB-tissue contact surface during freezing, which may increase the incidence of phrenic nerve (PN) palsy (PNP). OBJECTIVE: To evaluate the safety and efficacy of paroxysmal atrial fibrillation (AF) ablation with the ARC-Adv-CB as well as the merits of a predictor of PNP. METHODS: AF ablation was performed by using a "single 28-mm big CB" approach. The rate of pulmonary vein (PV) isolation with a first cryoapplication was measured. The distance between the CB and a PN pacing catheter in the superior vena cava was measured to predict PNP during freezing. RESULTS: In 147 patients, PV were isolated with a single cryoapplication in 205 (81.3%) of 252 PV treated with the ARC-CB and in 280 (90.3%) of 310 PV treated with the ARC-Adv-CB (P = .003). The mean time to PV isolation was 52 ± 34 seconds and 40 ± 25 seconds (P < .001) and the temperature at the time of isolation was -36.1 ± 10.3°C and -32.3 ± 10.2°C (P = .001) in the ARC-CB and ARC-Adv-CB groups, respectively. Mean procedure and fluoroscopy durations were significantly shorter in the ARC-Adv-CB group. Transient PNP was observed in 7(10.6%) and 20(24.4%) of the patients treated with the ARC-CB and ARC-Adv-CB, respectively (P = .048). The distance between the lateral edge of the CB and a vertical line through the tip of the pacing catheter accurately predicted PNP (P < .001). CONCLUSIONS: The 28-mm ARC-Adv-CB enabled more efficient ablation of paroxysmal AF and shorter procedures than did the ARC-CB. This higher performance was associated with a higher incidence of PNP, which was predicted by the distance between the CB and the PN. Article in Journal/Newspaper Human health Inserm: HAL (Institut national de la santé et de la recherche médicale) Heart Rhythm 11 3 386 393
institution Open Polar
collection Inserm: HAL (Institut national de la santé et de la recherche médicale)
op_collection_id ftinserm
language English
topic Cryoablation
Paroxysmal atrial fibrillation
Cryoballoon ablation
Pulmonary vein
Pulmonary vein isolation
Phrenic nerve palsy
[SDV.MHEP]Life Sciences [q-bio]/Human health and pathology
spellingShingle Cryoablation
Paroxysmal atrial fibrillation
Cryoballoon ablation
Pulmonary vein
Pulmonary vein isolation
Phrenic nerve palsy
[SDV.MHEP]Life Sciences [q-bio]/Human health and pathology
Martins, Raphaël
Hamon, David
Césari, Olivier
Behaghel, Albin
Behar, Nathalie
Sellal, Jean-Marc
Daubert, Jean-Claude
Mabo, Philippe
Pavin, Dominique
Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation
topic_facet Cryoablation
Paroxysmal atrial fibrillation
Cryoballoon ablation
Pulmonary vein
Pulmonary vein isolation
Phrenic nerve palsy
[SDV.MHEP]Life Sciences [q-bio]/Human health and pathology
description International audience BACKGROUND: Compared with the first-generation Arctic Front cryoballoon (ARC-CB), the new Arctic Front Advance cryoballoon (ARC-Adv-CB) increases the efficient CB-tissue contact surface during freezing, which may increase the incidence of phrenic nerve (PN) palsy (PNP). OBJECTIVE: To evaluate the safety and efficacy of paroxysmal atrial fibrillation (AF) ablation with the ARC-Adv-CB as well as the merits of a predictor of PNP. METHODS: AF ablation was performed by using a "single 28-mm big CB" approach. The rate of pulmonary vein (PV) isolation with a first cryoapplication was measured. The distance between the CB and a PN pacing catheter in the superior vena cava was measured to predict PNP during freezing. RESULTS: In 147 patients, PV were isolated with a single cryoapplication in 205 (81.3%) of 252 PV treated with the ARC-CB and in 280 (90.3%) of 310 PV treated with the ARC-Adv-CB (P = .003). The mean time to PV isolation was 52 ± 34 seconds and 40 ± 25 seconds (P < .001) and the temperature at the time of isolation was -36.1 ± 10.3°C and -32.3 ± 10.2°C (P = .001) in the ARC-CB and ARC-Adv-CB groups, respectively. Mean procedure and fluoroscopy durations were significantly shorter in the ARC-Adv-CB group. Transient PNP was observed in 7(10.6%) and 20(24.4%) of the patients treated with the ARC-CB and ARC-Adv-CB, respectively (P = .048). The distance between the lateral edge of the CB and a vertical line through the tip of the pacing catheter accurately predicted PNP (P < .001). CONCLUSIONS: The 28-mm ARC-Adv-CB enabled more efficient ablation of paroxysmal AF and shorter procedures than did the ARC-CB. This higher performance was associated with a higher incidence of PNP, which was predicted by the distance between the CB and the PN.
author2 Service de cardiologie et maladies vasculaires Rennes = Cardiac, Thoracic, and Vascular Surgery Rennes
Centre Hospitalier Universitaire de Rennes CHU Rennes = Rennes University Hospital Pontchaillou
Laboratoire Traitement du Signal et de l'Image (LTSI)
Université de Rennes (UR)-Institut National de la Santé et de la Recherche Médicale (INSERM)
CIC-IT Rennes
Centre Hospitalier Universitaire de Rennes CHU Rennes = Rennes University Hospital Pontchaillou -Institut National de la Santé et de la Recherche Médicale (INSERM)
Clinique Saint Gatien
format Article in Journal/Newspaper
author Martins, Raphaël
Hamon, David
Césari, Olivier
Behaghel, Albin
Behar, Nathalie
Sellal, Jean-Marc
Daubert, Jean-Claude
Mabo, Philippe
Pavin, Dominique
author_facet Martins, Raphaël
Hamon, David
Césari, Olivier
Behaghel, Albin
Behar, Nathalie
Sellal, Jean-Marc
Daubert, Jean-Claude
Mabo, Philippe
Pavin, Dominique
author_sort Martins, Raphaël
title Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation
title_short Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation
title_full Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation
title_fullStr Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation
title_full_unstemmed Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation
title_sort safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation
publisher HAL CCSD
publishDate 2014
url https://hal.univ-lorraine.fr/hal-03233143
https://doi.org/10.1016/j.hrthm.2014.01.002
genre Human health
genre_facet Human health
op_source ISSN: 1556-3871
EISSN: 1547-5271
Heart Rhythm
https://hal.univ-lorraine.fr/hal-03233143
Heart Rhythm, 2014, 11 (3), pp.386-393. &#x27E8;10.1016/j.hrthm.2014.01.002&#x27E9;
op_relation info:eu-repo/semantics/altIdentifier/doi/10.1016/j.hrthm.2014.01.002
info:eu-repo/semantics/altIdentifier/pmid/24389575
hal-03233143
https://hal.univ-lorraine.fr/hal-03233143
doi:10.1016/j.hrthm.2014.01.002
PUBMED: 24389575
op_doi https://doi.org/10.1016/j.hrthm.2014.01.002
container_title Heart Rhythm
container_volume 11
container_issue 3
container_start_page 386
op_container_end_page 393
_version_ 1810448721672929280