Reproducibility of TL Measurements in a Mixed Field of Thermal Neutrons and Photons

The reproducibility of measurements performed with GR-100 (LiF:Mg,Ti) from the Solid Dosimetric Detector and Method Laboratory (DML) China, GR-107 (7LiF:Mg,Ti, DML), TLD-700H (7LiF:Mg,Cu,P, Harshaw) and Al 2 O 3 :Mg,Y (Hungary) in photon and mixed photon-neutron fields was investigated. Mixed-field...

Full description

Bibliographic Details
Published in:Radiation Protection Dosimetry
Main Authors: C. Fernandes, A., C. Gonçalves, I., Ferro Carvalho, A., Santos, J., Cardoso, J., Santos, L., Osvay, M.
Format: Text
Language:English
Published: Oxford University Press 2002
Subjects:
DML
Online Access:http://rpd.oxfordjournals.org/cgi/content/short/101/1-4/481
https://doi.org/10.1093/oxfordjournals.rpd.a006032
Description
Summary:The reproducibility of measurements performed with GR-100 (LiF:Mg,Ti) from the Solid Dosimetric Detector and Method Laboratory (DML) China, GR-107 (7LiF:Mg,Ti, DML), TLD-700H (7LiF:Mg,Cu,P, Harshaw) and Al 2 O 3 :Mg,Y (Hungary) in photon and mixed photon-neutron fields was investigated. Mixed-field irradiations were performed in a thermal neutron field generated at a nuclear reactor. GR-100 sensitivity decreased after mixed-field irradiations, while no significant change was found for the other materials. Using GR-100 for the dosimetry of mixed and high-intensity fields requires careful procedures.