Simultaneous trace gas measurements using two Fourier transform spectrometers at Eureka, Canada during spring 2006, and comparisons with the ACE-FTS

The 2006 Canadian Arctic ACE (Atmospheric Chemistry Experiment) Validation Campaign collected measurements at the Polar Environment Atmospheric Research Laboratory (PEARL, 86.42° W, 80.05° N, 610 m a.s.l.) at Eureka, Canada from 17 February to 31 March 2006. Two of the ten instruments involved in th...

Full description

Bibliographic Details
Published in:Atmospheric Chemistry and Physics
Main Authors: Fu, D., Walker, K. A., Mittermeier, R. L., Strong, K., Sung, K., Fast, H., Bernath, P. F., Boone, C. D., Daffer, W. H., Fogal, P., Kolonjari, F., Loewen, P., Manney, G. L., Mikhailov, O., Drummond, J. R.
Format: Text
Language:English
Published: 2018
Subjects:
Online Access:https://doi.org/10.5194/acp-11-5383-2011
https://www.atmos-chem-phys.net/11/5383/2011/
id ftcopernicus:oai:publications.copernicus.org:acp5113
record_format openpolar
spelling ftcopernicus:oai:publications.copernicus.org:acp5113 2023-05-15T15:18:58+02:00 Simultaneous trace gas measurements using two Fourier transform spectrometers at Eureka, Canada during spring 2006, and comparisons with the ACE-FTS Fu, D. Walker, K. A. Mittermeier, R. L. Strong, K. Sung, K. Fast, H. Bernath, P. F. Boone, C. D. Daffer, W. H. Fogal, P. Kolonjari, F. Loewen, P. Manney, G. L. Mikhailov, O. Drummond, J. R. 2018-08-09 application/pdf https://doi.org/10.5194/acp-11-5383-2011 https://www.atmos-chem-phys.net/11/5383/2011/ eng eng doi:10.5194/acp-11-5383-2011 https://www.atmos-chem-phys.net/11/5383/2011/ eISSN: 1680-7324 Text 2018 ftcopernicus https://doi.org/10.5194/acp-11-5383-2011 2019-12-24T09:56:50Z The 2006 Canadian Arctic ACE (Atmospheric Chemistry Experiment) Validation Campaign collected measurements at the Polar Environment Atmospheric Research Laboratory (PEARL, 86.42° W, 80.05° N, 610 m a.s.l.) at Eureka, Canada from 17 February to 31 March 2006. Two of the ten instruments involved in the campaign, both Fourier transform spectrometers (FTSs), were operated simultaneously, recording atmospheric solar absorption spectra. The first instrument was an ABB Bomem DA8 high-resolution infrared FTS. The second instrument was the Portable Atmospheric Research Interferometric Spectrometer for the Infrared (PARIS-IR), the ground-based version of the satellite-borne FTS on the ACE satellite (ACE-FTS). From the measurements collected by these two ground-based instruments, total column densities of seven stratospheric trace gases (O 3 , HCl, ClONO 2 , HF, HNO 3 , NO 2 , and NO) were retrieved using the optimal estimation method and these results were compared. Since the two instruments sampled the same portions of atmosphere by synchronizing observations during the campaign and used consistent retrieval parameters, the biases in retrieved columns from the two spectrometers represent the instrumental differences. Mean differences in total column densities of O 3 , HCl, ClONO 2 , HF, HNO 3 , and NO 2 from the observations between PARIS-IR and the DA8 FTS are 2.8 %, −3.2 %, −4.3 %, −1.5 %, −1.9 %, and −0.1 %, respectively. Partial column results from the ground-based spectrometers were also compared with partial columns derived from ACE-FTS version 2.2 (including updates for O 3 ) profiles. Mean differences in partial column densities of O 3 , HCl, ClONO 2 , HF, HNO 3 , NO 2 , and NO from the measurements between ACE-FTS and the DA8 FTS are −5.9 %, −8.5 %, −11.8 %, −0.9 %, −6.6 %, −21.6 % and −7.6 % respectively. Mean differences in partial column densities of O 3 , HCl, ClONO 2 , HF, HNO 3 , NO 2 from the measurements between ACE-FTS and the PARIS-IR are −5.2 %, −4.6 %, −2.3 %, −4.7 %, 5.7 % and −11.9 %, respectively. This work provides further evidence of the reliability of ACE-FTS measurements from the first three years of on-orbit observations. Column densities of O 3 , HCl, ClONO 2 , and HNO 3 from the three FTSs were normalized with respect to HF and used to compare the time evolution of the chemical constituents in the atmosphere over Eureka during spring 2006. Text Arctic Copernicus Publications: E-Journals Arctic Canada Eureka ENVELOPE(-85.940,-85.940,79.990,79.990) Atmospheric Chemistry and Physics 11 11 5383 5405
institution Open Polar
collection Copernicus Publications: E-Journals
op_collection_id ftcopernicus
language English
description The 2006 Canadian Arctic ACE (Atmospheric Chemistry Experiment) Validation Campaign collected measurements at the Polar Environment Atmospheric Research Laboratory (PEARL, 86.42° W, 80.05° N, 610 m a.s.l.) at Eureka, Canada from 17 February to 31 March 2006. Two of the ten instruments involved in the campaign, both Fourier transform spectrometers (FTSs), were operated simultaneously, recording atmospheric solar absorption spectra. The first instrument was an ABB Bomem DA8 high-resolution infrared FTS. The second instrument was the Portable Atmospheric Research Interferometric Spectrometer for the Infrared (PARIS-IR), the ground-based version of the satellite-borne FTS on the ACE satellite (ACE-FTS). From the measurements collected by these two ground-based instruments, total column densities of seven stratospheric trace gases (O 3 , HCl, ClONO 2 , HF, HNO 3 , NO 2 , and NO) were retrieved using the optimal estimation method and these results were compared. Since the two instruments sampled the same portions of atmosphere by synchronizing observations during the campaign and used consistent retrieval parameters, the biases in retrieved columns from the two spectrometers represent the instrumental differences. Mean differences in total column densities of O 3 , HCl, ClONO 2 , HF, HNO 3 , and NO 2 from the observations between PARIS-IR and the DA8 FTS are 2.8 %, −3.2 %, −4.3 %, −1.5 %, −1.9 %, and −0.1 %, respectively. Partial column results from the ground-based spectrometers were also compared with partial columns derived from ACE-FTS version 2.2 (including updates for O 3 ) profiles. Mean differences in partial column densities of O 3 , HCl, ClONO 2 , HF, HNO 3 , NO 2 , and NO from the measurements between ACE-FTS and the DA8 FTS are −5.9 %, −8.5 %, −11.8 %, −0.9 %, −6.6 %, −21.6 % and −7.6 % respectively. Mean differences in partial column densities of O 3 , HCl, ClONO 2 , HF, HNO 3 , NO 2 from the measurements between ACE-FTS and the PARIS-IR are −5.2 %, −4.6 %, −2.3 %, −4.7 %, 5.7 % and −11.9 %, respectively. This work provides further evidence of the reliability of ACE-FTS measurements from the first three years of on-orbit observations. Column densities of O 3 , HCl, ClONO 2 , and HNO 3 from the three FTSs were normalized with respect to HF and used to compare the time evolution of the chemical constituents in the atmosphere over Eureka during spring 2006.
format Text
author Fu, D.
Walker, K. A.
Mittermeier, R. L.
Strong, K.
Sung, K.
Fast, H.
Bernath, P. F.
Boone, C. D.
Daffer, W. H.
Fogal, P.
Kolonjari, F.
Loewen, P.
Manney, G. L.
Mikhailov, O.
Drummond, J. R.
spellingShingle Fu, D.
Walker, K. A.
Mittermeier, R. L.
Strong, K.
Sung, K.
Fast, H.
Bernath, P. F.
Boone, C. D.
Daffer, W. H.
Fogal, P.
Kolonjari, F.
Loewen, P.
Manney, G. L.
Mikhailov, O.
Drummond, J. R.
Simultaneous trace gas measurements using two Fourier transform spectrometers at Eureka, Canada during spring 2006, and comparisons with the ACE-FTS
author_facet Fu, D.
Walker, K. A.
Mittermeier, R. L.
Strong, K.
Sung, K.
Fast, H.
Bernath, P. F.
Boone, C. D.
Daffer, W. H.
Fogal, P.
Kolonjari, F.
Loewen, P.
Manney, G. L.
Mikhailov, O.
Drummond, J. R.
author_sort Fu, D.
title Simultaneous trace gas measurements using two Fourier transform spectrometers at Eureka, Canada during spring 2006, and comparisons with the ACE-FTS
title_short Simultaneous trace gas measurements using two Fourier transform spectrometers at Eureka, Canada during spring 2006, and comparisons with the ACE-FTS
title_full Simultaneous trace gas measurements using two Fourier transform spectrometers at Eureka, Canada during spring 2006, and comparisons with the ACE-FTS
title_fullStr Simultaneous trace gas measurements using two Fourier transform spectrometers at Eureka, Canada during spring 2006, and comparisons with the ACE-FTS
title_full_unstemmed Simultaneous trace gas measurements using two Fourier transform spectrometers at Eureka, Canada during spring 2006, and comparisons with the ACE-FTS
title_sort simultaneous trace gas measurements using two fourier transform spectrometers at eureka, canada during spring 2006, and comparisons with the ace-fts
publishDate 2018
url https://doi.org/10.5194/acp-11-5383-2011
https://www.atmos-chem-phys.net/11/5383/2011/
long_lat ENVELOPE(-85.940,-85.940,79.990,79.990)
geographic Arctic
Canada
Eureka
geographic_facet Arctic
Canada
Eureka
genre Arctic
genre_facet Arctic
op_source eISSN: 1680-7324
op_relation doi:10.5194/acp-11-5383-2011
https://www.atmos-chem-phys.net/11/5383/2011/
op_doi https://doi.org/10.5194/acp-11-5383-2011
container_title Atmospheric Chemistry and Physics
container_volume 11
container_issue 11
container_start_page 5383
op_container_end_page 5405
_version_ 1766349128100478976