Validity of a smartwatch for detecting atrial fibrillation in patients after heart valve surgery:a prospective observational study

Objectives Atrial fibrillation (AF) is a common early arrhythmia after heart valve surgery that limits physical activity. We aimed to evaluate the criterion validity of the Apple Watch Series 5 single-lead electrocardiogram (ECG) for detecting AF in patients after heart valve surgery. Design We enro...

Full description

Bibliographic Details
Published in:Scandinavian Cardiovascular Journal
Main Authors: Müller, Margrethe, Hanssen, Tove Aminda, Johansen, David, Jakobsen, Øyvind, Pedersen, John Erling, Aamot Aksetøy, Inger Lise, Rasmussen, Trine Bernholdt, Hartvigsen, Gunnar, Skogen, Vegard, Thrane, Gyrd
Format: Article in Journal/Newspaper
Language:English
Published: 2024
Subjects:
Online Access:https://curis.ku.dk/portal/da/publications/validity-of-a-smartwatch-for-detecting-atrial-fibrillation-in-patients-after-heart-valve-surgery(0573e44b-c5f6-4cc1-8bd0-bdc3f6945457).html
https://doi.org/10.1080/14017431.2024.2353069
https://curis.ku.dk/ws/files/394439545/Validity_of_a_smartwatch_for_detecting_atrial_fibrillation_in_patients_after_heart_valve_surgery_a_prospective_observational_study.pdf
Description
Summary:Objectives Atrial fibrillation (AF) is a common early arrhythmia after heart valve surgery that limits physical activity. We aimed to evaluate the criterion validity of the Apple Watch Series 5 single-lead electrocardiogram (ECG) for detecting AF in patients after heart valve surgery. Design We enrolled 105 patients from the University Hospital of North Norway, of whom 93 completed the study. All patients underwent single-lead ECG using the smartwatch three times or more daily on the second to third or third to fourth postoperative day. These results were compared with continuous 2–4 days ECG telemetry monitoring and a 12-lead ECG on the third postoperative day. Results On comparing the Apple Watch ECGs with the ECG monitoring, the sensitivity and specificity to detect AF were 91% (75, 100) and 96% (91, 99), respectively. The accuracy was 95% (91, 99). On comparing Apple Watch ECG with a 12-lead ECG, the sensitivity was 71% (62, 100) and the specificity was 92% (92, 100). Conclusion The Apple smartwatch single-lead ECG has high sensitivity and specificity, and might be a useful tool for detecting AF in patients after heart valve surgery. Objectives: Atrial fibrillation (AF) is a common early arrhythmia after heart valve surgery that limits physical activity. We aimed to evaluate the criterion validity of the Apple Watch Series 5 single-lead electrocardiogram (ECG) for detecting AF in patients after heart valve surgery. Design: We enrolled 105 patients from the University Hospital of North Norway, of whom 93 completed the study. All patients underwent single-lead ECG using the smartwatch three times or more daily on the second to third or third to fourth postoperative day. These results were compared with continuous 2–4 days ECG telemetry monitoring and a 12-lead ECG on the third postoperative day. Results: On comparing the Apple Watch ECGs with the ECG monitoring, the sensitivity and specificity to detect AF were 91% (75, 100) and 96% (91, 99), respectively. The accuracy was 95% (91, 99). On comparing Apple ...